back to the overview
Tamson Low Temperature Circulator from -80°C up to 20°C.
Article number: 00T0540
Specification:
file_icon Specification TLC80DP file_icon Manual TLC80DP file_icon PPT TLC80-DP

Remarks:

The TLC80-14DP is a low temperature circulator with a bath contents of 14 litres. The minimum temperature which can be reached is minus 82°C. At -80°C the heat removal capacity is 150 Watts. The bath can be used for general low temperature use but shows excellent heat removal performance i.e. in combination with Cloud and Pour Point tests. Low fluid level is detected electronically. During the nights the TLC80-14DP can be switched to economical mode. The device is especially developed to cool down 2 or more apparatuses. The manifold can be used for this application.

The cooling system inside the TLC 80 is ozone friendly and doesn't contain any CFK/HCFK gas.

The TLC80-DP combines a wide temperature range and  maintenance free -long life- cooling sys­tem. Accurate temperature regulation is provided by a microprocessor controlled PID system also enabling an "autotune" function for optimum setting of the temperature regulation.

The TLC80 operates from -80°C to +20°C (-40°F..68°F).

The set point can be set in steps of 0.1°K. The overall system accuracy is ±0.05°K. Temperature readout in 0.1°K.

A pump provides circulation in the bath or via an external circuit. The pump offers 10 litres per minute maximum pressure is 300mBar. Optional pump with pressure max 1bar, flow max 16ltrs/min, reference: 24T0392.

The bath conforms to CE regulations.

Available in 230V/50Hz.


Standards:

back to the overview

TLC80-14DP ( Lowtemp baths and Circulators)