back to the overview
Tamson Low Temperature Circulator from -40°C up to 20°C.
Article number: 00T0520
Specification:
file_icon Specification TLC40 file_icon Manual TLC40 file_icon PPT TLC40

Remarks:

The TLC40 cooling circulator can be used for multiple heat removal purposes such as replacement of tapwater, fast cool down, distillation or ASTM CFPP testing. The bath content is 14 litres.

The cooling system inside the TLC40 is ozone friendly and doesn't contain any CFK/HCFK gas.

The TLC40 combines a wide temperature range and  maintenance free -long life- cooling sys­tem. Accurate temperature regulation is provided by a microprocessor controlled PID system also enabling an "autotune" function for optimum setting of the temperature regulation. The cooling system provides rapid cooling down of the 14 litres bath contents.

The TLC40 operates from -40°C to +20°C (-40°F..68°F) and optional extended minimum temperature:  -45°C .. +20°C / -49°F .. + 68°F.

The setpoint can be set in steps of 0.1°K. The overall system accuracy is ±0.05°K. Temperature readout in 0.1°K

A pump provides circulation in the bath or via an external circuit. The pump offers 7 litres per minute maximum pressure is 300mBar. Optional pump with pressure max 1bar, flow max 16ltrs/min, reference: 24T0392

The bath conforms to CE regulations.

Available in 230V/50Hz, 230V/60Hz and 115V/60Hz.


Standards:

back to the overview

TLC40-14 ( Lowtemp baths and Circulators)

Nederlands | English